1.3 Required tools and components
You will need the following tools and components to install the 700XA
Zero grade carrier gas:
99.995% pure
Less than 5 ppm water
Less than 0.5 ppm hydrocarbons
High pressure dual-stage regulator for the carrier gas cylinder
High side up to 3000 psig (20684.3 kPa)
Gauge (psig)
Low side capable of controlling pressure up to 150 psig (1034.2 kPa)
Calibration standard gas with correct number of components and concentrations
Dual-stage regulator for the calibration gas cylinder with a low pressure side capable of
controlling pressure up to 30 psig (206.9 kPa)
Sample probe regulator (fixture for procuring the stream or sample gas for
chromatographic analysis)
Coalescing filter
Membrane filter
-in. stainless steel tubing
For connecting calibration gas to the GC
For connecting carrier gas to the GC
For connecting stream gas to the GC
Sulfinert tubing required if sulfur components are present in calibration gas
Heat tracing, as required for sample transport and calibration lines
Miscellaneous tube fittings, tubing benders, and tubing cutter
14 American wire gauage (AWG) (18 metric wire gauge [MWG]) or larger electrical
wiring and conduit to provide 120 or 240 Vac, single phase, 50 to 60 Hz, from an
appropriate circuit breaker and power disconnect switch. See guidelines in Gas
chromatograph wiring.
Digital volt-ohm meter with probe-type leads
Flow measuring device
Open-end wrenches sized ¼-in., 5/16-in., 7/16-in., ½-in., 9/16-in., and -in.
Torque wrench
Quick Start Guide Getting started
00825-0100-3770 July 2020
Rosemount 700XA 11