b) The RJ45 Ethernet port at J22: Enter the appropriate values in the Ethernet1
IP Address, the Ethernet1 Subnet, and the Default Gateway fields.
Note
See your information technology (IT) staff to obtain IP, subnet, and gateway
addresses.
Important
To configure a Ethernet IP address using the local operator interface (LOI),
refer to Figure 9.
4. Click OK.
5. Log off the GC.
6. Access the backplane, which is located in the GC’s enclosure.
7. If you are setting up a static IP address for the Ethernet1 port at J22, and you also
intend to connect to your company’s local area network, do the following:
a) Locate , at SW1 on the backplane.
SW1 is located the Ethernet port at J22 .
b)
This disables the DHCP server.
8. To connect to the GC, do the following:
a) Start MON2020 and select File GC Directory....
The GC Directory window displays.
b) Select Add.
MON2020 adds a new GC profile to the end of the table.
Note
You can name the GC’s profile as well as add a short description.
c) Select the new profile and click Ethernet... Enter the GC’s static IP address in
the IP address field.
d) Click OK.
The Ethernet Connection Properties for New GC window closes.
9. Click Save to save the new profile.
10. Click OK to close the GC Directory window.
11. Select Chromatograph Connect... to connect to the GC or click
.
The Connect to GC window displays. The newly created GC profile should be listed in
the table.
12. Locate the new GC profile and click the Ethernet button that is associated with it.
The Login window displays.
13. Enter a User Name and User Pin and click OK.
Installation and start-up Quick Start Guide