INSTALLATION AND GROUNDING INSTRUCTIONS
Examine Your Warmer
Unpack warmer, remove all packing material and examine the unit for any damage such as dents, broken door hinges or cracks in the door. Notify Sage Products
immediately if warmer is damaged. DO NOT install or operate if warmer is damaged.
Remove Protective Plastic Film
If a plastic film has been wrapped around the warmer to protect the painted finish from scratching during shipping, please remove this plastic film prior to use.
(However, not all warmers have plastic film wrapping.)
Attach Power Cord to Warmer
Power cord ships inside warmer; remove from inside unit and attach to Warmer.
Placement of Warmer
1. Warmer must be placed on a flat, stable surface. For proper operation, the warmer must have sufficient air flow. Allow 3 inches (7.62 cm) of space on both sides
of the warmer and 1 inch (2.54 cm) of space on top of the warmer.
a) DO NOT block air vent. If it is blocked during operation, the warmer may overheat. If overheating occurs, a thermal safety device will turn the unit off.
The warmer will remain inoperable until it has cooled.
b) DO NOT place warmer near a hot, damp surface, such as a gas or electric range.
c) Be certain that power cord on the warmer can reach the electrical outlet.
d) Be certain electrical outlet is not blocked by the warmer. Power cord must be able to be unplugged without moving the warmer.
2. Warmer should be located in or near the nurse’s station or supply room, and at least 5 feet (1.5 m) from patients.
Grounding Instructions
This appliance must be grounded. In the event of an electrical short circuit, grounding reduces the risk of an electric shock by providing an escape wire for the
electric current. This appliance is equipped with a cord having a grounded wire with a grounded plug. The plug must be plugged into an outlet that is properly
installed and grounded.
WARNING: Improper use of the grounding plug can result in a risk of electric shock.
Consult a qualified electrician or service person if the grounding instructions are not completely understood, or if doubt exists as to whether the appliance is
properly grounded.
If it is necessary to use an extension cord, use only a three slot receptacle that will accept the plug on the appliance. The marked rating of the extension cord should
be equal to or greater than the electrical rating of the appliance.
Wiring Requirements
1. US Configuration:
Rating: 120V AC / 60 Hz / 7 AMP input.
EU Configuration:
Rating: 240V AC / 50 Hz / 3.5 AMP input.
2. The power cord is to be used for mains disconnection.
3. For US configuration, a hospital-grade receptacle is required.
4. For US customers without a proper hospital-grade receptacle, it is the responsibility and obligation of the user to have it replaced with a properly grounded,
hospital-grade receptacle.
5. Using incorrect voltage may result in fire or other accident causing warmer damage. Sage Products LLC is not responsible for damage resulting from use of the
warmer with voltages other than those specified. Voltage used must be the same as that specified on equipment rating label.
6. The warmer must be on a SEPARATE CIRCUIT. No other appliance, with the exception of another Sage Warmer, should share the circuit with the warmer. If it
does, the branch circuit fuse may blow, the circuit breaker may trip or products may warm slower than expected.
7. If using more than one Sage Warmer: A maximum of 2 Sage Warmers may be used on the same circuit. Do not exceed capacity of circuit.
- 4 -