E230169
Medical equipment with respect to electric shock, fire and mechanical
hazards only, in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1.
Consult Instructions for Use
Protective Earth (Ground)
TECHNICAL SPECIFICATIONS
Product Code 7937 (US Configuration) 7937-X (EU configuration)
Input Rating 120 V AC / 60 Hz / 7 AMP 240 V AC / 50 Hz / 3.5 AMP
Target Temperature (B Configuration) Adjustable between 100°F and 125°F (38°C - 52°C) Adjustable between 100°F and 125°F (38°C - 52°C)
Target Temperature (C Configuration) Adjustable between 100°F and 125°F (38°C - 52°C) Adjustable between 100°F and 125°F (38°C - 52°C)
Default Temperature Setting
(B Configuration)
125°F (52°C) 125°F (52°C)
Default Temperature Setting
(C Configuration)
125°F (52°C) 125°F (52°C)
Ambient Operating Range 60°F - 80°F (16°C - 27°C) 60°F - 80°F (16°C - 27°C)
Relative Humidity Operating Range 50 - 70% 50 - 70%
Operating Pressure 70 kPa - 106 kPa 70 kPa - 106 kPa
Outer Dimensions (H x W x D) 16.88” x 14.69” x 15.25” (429mm x 373mm x 387mm) 16.88” x 14.69” x 15.25” (429mm x 373mm x 387mm)
Net Weight (N.W.)
Gross Weight (G.W.)
N.W.: 29.5 lb. (13.4 kg)
G.W.: 34 lb. (15.5 kg)
N.W.: 29.5 lb. (13.4 kg)
G.W.: 34 lb. (15.5 kg)
Mass with Max Working Load
(fully loaded unit)
36.8 lb. (16.7 kg) 36.8 lb. (16.7 kg)
Mass with Max Working Load
(fully loaded cart and unit)
117.3 lb. (53.3 kg) 117.3 lb. (53.3 kg)
Classification Class I Class I
Degree of Protection Against Ingress
of Water
IPXO IPXO
Mode of Operation Suitable for Continuous Operation Suitable for Continuous Operation
Storage / Transport Temperature -22 - 149°F (-30 - 65°C) -22 - 149°F (-30 - 65°C)
Storage / Transport Relative Humidity 10% to 90% 10% to 90%
Storage / Transport Atmospheric
Pressure Range
500 to 1060 hPa 500 to 1060 hPa
Standards - UL / EN / CSA / IEC 60601-1 UL 60601-1 Medical Electrical Equipment UL 60601-1 Medical Electrical Equipment
UL Classification IEC 60601-1, 2nd Edition IEC 60601-1, 2nd Edition
Third Party Evaluation and Testing IEC 60601-1, 3rd Edition IEC 60601-1, 3rd Edition
Characteristics
Powder Coated Steel Exterior / Plastic Door and Window /
Steel Front Flange / Aluminum Interior and Shelves
Powder Coated Steel Exterior / Plastic Door and Window /
Steel Front Flange / Aluminum Interior and Shelves
Fuse Type T10AH, 250V T10AH, 250V
Disposal Follow national requirements Follow national requirements
Refer to accompanying documents Refer to accompanying documents
Protective earth ground terminal Protective earth ground terminal
Applied Parts No applied parts No applied parts
Caution
Equipment not suitable for use in the presence of
flammable anesthetics
Equipment not suitable for use in the presence of
flammable anesthetics
- 7 -