Tissue-Tek® Glas™ Coverslipper
Quick Reference Guide
Tissue-Tek® Glas™ Coverslipper
page 6 | 6 QRG | Glas Coverslipper | 2011 | V-2
6 Months
Anti-drying bottle
1. Change the solvent for mounting medium* in
the anti-drying bottle.
Cleaning the receiving racks
1. Soak the racks for 24 hours in solvent for
mounting medium*.
Instrument strainer
Open cover on the right side of the instrument and
remove the strainer cap. Use xylene soaked brush
to clean the filter. Replace the cap.
Yearly
1. Preventive maintenance by a Sakura
representative or a qualified service engineer.
Trouble shooting
(section 6 operating manual)
The trouble shooting part of the operating manual
is divided into three sections to explain how to
recover from conditions that may occur during
routine operation.
In the first section,
Troubleshooting Chart is provided, which identifies
possible conditions related to the electrical and
mechanical operations.
The second section,
Messages, explains specific messages that may be
displayed during routine operation in detail.
The third section,
Gives a detailed explanation of all error codes and
how to recover from them.
Operating manual section 6, trouble shooting
1. Change the solvent for mounting medium* in
the loading station.
* Solvent for mounting medium is the reagent which is compatible
with the mounting medium, Tissue clear when using Tissue Mount
and Xylene when using Glass coverslipping medium.
Waste container, spatula part