Samsung GM85 Quick User Guide.
(Please note that this guide is aimed at enabling a user to perform a basic mobile chest x-ray
and is in no way a substitute for full training which includes safety, detector care, cleaning,
advanced functions, errors and problem solving)
Turning on
1. If the unit is off, switch on with the power switch on top (next to the emergency stop
button).
2. Unplug the system from the mains and enter the drive pin to enable movement.
3. Log-in to the S-Station to use the unit and release the detector locks.
Worklist Screen
1. Connect to the network via a cable and obtain the worklist. Prompt the system by
pressing refresh (circular arrow, bottom left hand of screen).
2. Disconnect from network and head to patient.
3. Once by the patient, lift the tube, select patient and press ‘Start Exam
Acquisition Screen
1. Make sure the detector is turned on (power button on the side edge of detector)
2. Position as normal
3. Check exposure factors and expose.
Review Screen
1. Check the image
- If okay, retrieve the detector.
- If repeat required, press ‘RETAKE’, select reason for repeat and press okay.
2. Shutter, annotate and window image as appropriate.
3. Once happy with the image press ‘CONFIRM’ to save changes to the image and then
press ‘AUTO SEND’ to close the patient and send images to PACS.
4. Once back in the department, re-connect the unit to the network cable, images in
queue will now send
5. Check status of image Y/N in review tab and image has arrived on PACS.
Turning Off
1. The mobile unit should always be left plugged in and charging.
2. Press and hold power button for 3 seconds until shutdown noise heard
3. Lock detectors.
Question and Answer IconNeed help?

Do you have a question about the Samsung GM85 and is the answer not in the manual?

Samsung GM85 Specifications

General IconGeneral
Network TechnologyGSM / HSPA / LTE
SIMSingle SIM (Nano-SIM) or Dual SIM (Nano-SIM, dual stand-by)
3.5mm jackYes
StatusAvailable. Released 2018, March
ProtectionCorning Gorilla Glass 5
Memory64GB 4GB RAM
Camera FeaturesLED flash, panorama, HDR
Selfie Camera8 MP, f/2.0
LoudspeakerYes
WLANWi-Fi 802.11 b/g/n, Wi-Fi Direct
Bluetooth5.0, A2DP, LE
GPSYes, with A-GPS, GLONASS, GALILEO, BDS
RadioFM radio
USBUSB Type-C
SensorsFingerprint (side-mounted), accelerometer, proximity
ChargingFast charging 15W

Summary

Turning On

Power On Procedure

Switch on the unit using the power switch located on top, adjacent to the emergency stop button.

Prepare for Movement

Disconnect the system from mains power and insert the drive pin to enable unit mobility.

Log in and Release Locks

Log in to the S-Station to operate the unit and unlock the detector.

Worklist Screen Operations

Connect and Obtain Worklist

Connect to the network, obtain the worklist, and refresh the system display.

Disconnect and Proceed to Patient

Disconnect from the network and move the unit to the patient's location.

Select Patient and Start Exam

Position the unit by the patient, select the patient, and initiate the examination.

Acquisition Screen Steps

Detector Power Check

Ensure the detector is powered on using its side edge power button.

Patient Positioning

Position the patient for the X-ray examination as per standard procedure.

Set Exposure and Expose

Verify exposure factors and perform the X-ray exposure.

Review Screen Workflow

Image Review and Repeat

Check the acquired image; retrieve the detector if satisfactory or retake if needed.

Image Adjustments

Apply shutter, annotation, and windowing adjustments to the image as necessary.

Confirm and Send Image

Confirm image changes, close the patient record, and send images to PACS.

Reconnect and Send Queued Images

Reconnect to the network cable to send any queued images.

Verify Image Transmission

Check the review tab for image transmission status and PACS arrival.

Turning Off Procedures

Charging Recommendation

Ensure the mobile unit remains plugged in for charging.

Initiate Shutdown

Press and hold the power button for 3 seconds to initiate the shutdown sequence.

Lock Detectors

Secure the detectors by locking them in place.