EasyManua.ls Logo

Sanyo MCO-19MUVH - Page 76

Sanyo MCO-19MUVH
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
75
Name Multi-gas Incubator
Models MCO-19M MCO-19M(UV) MCO-19M(UVH)
Accessories
3 trays, 3 sets of tray supports, 3 gas tubes, 1 humidifying pan,
1 gas injection nozzle, 1 gas injection nozzle tube
1 pair of stacking plates A and B, 2 protective stickers, 6 tube bands
Optional accessories
H
2
O
2
Generator (MCO-HP)
H
2
O
2
Decontamination Kit (MCO-HL)
Standard equipment
UV Lamp Add-on Kit
(MCO-19UVS)
Standard equipment
H
2
O
2
Decontamination Reagent (MCO-H2O2),
Automatic CO
2
Standard Gas Calibration Kit (MCO-SG),
Gas Pressure Regulator (MCO-100L),
Extra Trays (MCO-47ST, includes 2 tray supports),
Half Trays (MCO-25ST), Automatic CO
2
Cylinder Changeover System (MCO-21GC),
Roller Base (MCO-18RB), UV lamp replacement kit (MCO-20UV)
Stacking spacer (MCO-18SB, when stacking on the MCO-175),
Stacking spacer (MCO-21SB, when stacking on the MCO-20AIC)
Ethernet (LAN) Interface (MTR-L03), RS232C/RS485 Interface (MTR-480)
SANYO Data acquisition software (MTR-5000)
4-20mA Interface (MCO-420MA, USA only)
Note: Designs and specifications are subject to change without notice.
PERFORMANCE
Temperature control range
Ambient temperature+5 to 50
o
C (ambient temperature: 5 to 35
o
C)
Temperature distribution
± 0.25
o
C (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load)
Temperature variation
± 0.1
o
C (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load)
CO
2
control range
0 to 20%
CO
2
variation
± 0.15% (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load)
O
2
control range
1 to 18% and 22 to 80%
O
2
variation
± 0.2% (ambient temperature: 25
o
C, setting: 37
o
C, CO
2
: 5%, O
2
: 5%, no load)
Chamber humidity
95% ± 5%
Applicable environment
condition
Temperature: 5 to 35
o
C, Humidity: 80% max.
(The designed performance may not be obtained
If the ambient temperature is equal or less than 15
o
C)
Noise level
30 dB (A scale)
Power consumption
Max. 354 W Max. 335 W
Heat emission Max. 1,274 kJ/h
Max. 1,206 kJ/h
Rated voltage, frequency
110 to 120 V, 60 Hz 220 V, 60 Hz 220 to 240 V, 50 Hz
Amperage
Max. 3.1 A Max. 1.5 A Max. 1.5 A
Note: The unit with CE mark complies with EC directives.
Based on our measuring method.
Designs and specifications are subject to change without notice.
Get other manuals https://www.bkmanuals.com

Table of Contents

Related product manuals