EasyManua.ls Logo

Sanyo MCO-19MUVH - User Manual

Sanyo MCO-19MUVH
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
MCO-19M(UVH)
MCO-19M(UV)
MCO-19M
Multi Gas Incubator
INSTRUCTION MANUAL
Get other manuals https://www.bkmanuals.com

Table of Contents

Questions and Answers

  • M
    Michael KerrSep 12, 2025
    Why is the gas concentration slow to recover in Sanyo Laboratory Equipment?
    • V
      Valerie SolisSep 12, 2025
      If the gas concentration is slow to recover in your Sanyo Laboratory Equipment, consider these factors: the HEPA filter used for the incubator gas piping might be clogged (if the CO gas concentration is normal), there might be little gas remaining in the gas cylinder, the secondary pressure for the CO gas pressure regulator might not be at the specified value of 0.03 MPaG (0.3 kgf/cm2G, 4.3 psiG), the secondary pressure for the N /CO gas pressure regulator might not be at the specified value of 0.05 MPaG (0.5 kgf/cm2G, 7.1 psiG), or the gas tube might be blocked.
  • J
    Jeffery BentonAug 25, 2025
    What to do if Sanyo Medical Equipment Incubator does not operate at all?
    • C
      cassandra66Aug 25, 2025
      First, verify that the Incubator is properly plugged in. Then, check for any power outages or tripped circuit breakers that may be interrupting the power supply.
  • L
    lisa33Aug 26, 2025
    Why is a large quantity of gas being consumed by my Sanyo Medical Equipment?
    • D
      deborahbrownAug 27, 2025
      Check if the doors are being frequently opened and closed. Inspect gas connectors for leaks due to gas tube deterioration or pinhole leaks; consider replacing the gas tube annually. Also, check the packing seals for the gastight split doors and inner door for defects, and ensure the access hole is closed.
  • J
    Jeremy FullerAug 27, 2025
    What to do if the gas concentration does not match the set value in Sanyo MCO-19MUVH Medical Equipment?
    • M
      michaelbarnesAug 27, 2025
      Verify that the secondary pressure for the CO2 gas pressure regulator is at the specified value of 0.03 MPaG (0.3 kgf/cm2G, 4.3 psiG) and the secondary pressure for the N2 gas pressure regulator is at the specified value of 0.05 MPaG (0.5 kgf/cm2G, 7.1 psiG). Also, check if the gas tube is blocked.
  • N
    Nicole RamseyAug 29, 2025
    What to do if H2O2 decontamination cannot be performed on Sanyo MCO-19MUVH Medical Equipment?
    • P
      Paula FlemingAug 29, 2025
      If using the MCO-19M(UV)/19M, check if the optional MCO-HL is installed. If using the MCO-19M, confirm that both the optional MCO-HL and MCO-19UVS are installed. Verify that the UV lamp is not burned out, as this prevents H2O2 decontamination. Ensure the H2O2 Generator cable is properly connected, that the entire bottle of H2O2 Decontamination Reagent has not been used, and that the upper limit temperature alarm is not set to 50°C or higher.
  • C
    Crystal CarterAug 31, 2025
    How to unlock the outer door of Sanyo MCO-19MUVH?
    • L
      Lori FisherAug 31, 2025
      When the power is OFF, the electric lock engages, preventing the outer door from opening. Turn the power ON or use the accessory unlock key to override the electric lock. Note that the outer door is also electrically locked during decontamination.
  • K
    Kenneth SanchezNov 20, 2025
    What to do if an alarm has occurred in Sanyo Medical Equipment?
    • Y
      yhamptonNov 20, 2025
      When starting operation, ensure that the chamber temperature and CO2 gas concentration match the set values. Check the CO2 gas pressure regulator's secondary pressure is at 0.03 MPaG (0.3 kgf/cm2G. 4.3 psiG) and the N2 gas pressure regulator's secondary pressure is at 0.05 MPaG (0.5 kgf/cm2G, 7.1 psiG). Verify the gas tube is properly connected. During operation, ensure the upper limit alarm temperature is set at least 1°C higher than the chamber set temperature. Check if the temperature setting has been changed, the door has been left open, or a low-temperature object has been placed inside. If any of these are the case, the alarm will automatically clear if you wait. Also, check for gas leaks, changes to the CO2 gas concentration setting, and whether the gas cylinder is empty. Check the ...
  • D
    danielgarciaSep 14, 2025
    What to do if the outer door does not open on Sanyo Laboratory Equipment?
    • A
      Angela BaileySep 14, 2025
      If the outer door of your Sanyo Laboratory Equipment does not open, it could be due to the electric lock being engaged when the power is OFF. Turn the power ON or use the accessory unlock key to override the electric lock. Also, during decontamination, the outer door is electrically locked and will not open.
  • M
    Mr. Dale Johnson PhDSep 12, 2025
    Why are normal cultures not possible, and the gas concentration is suspect in Sanyo Laboratory Equipment?
    • M
      Melissa RogersSep 12, 2025
      If normal cultures are not possible and you suspect the gas concentration in your Sanyo Laboratory Equipment, check if the ambient air environment around the incubator is normal and if there is a source of polluted gas nearby.
  • H
    hwilsonSep 7, 2025
    Why is a large quantity of gas being consumed by my Sanyo Laboratory Equipment?
    • J
      Justin MckenzieSep 7, 2025
      If your Sanyo Laboratory Equipment incubator is consuming a large quantity of gas, consider the following: Are the doors being opened and closed frequently? Check for gas leaks from connectors due to deterioration of the gas tube, or any pinhole leaks. Inspect the packing seal for gastight split doors and the inner door for defects. Also, ensure the access hole is closed.

Sanyo MCO-19MUVH Specifications

General IconGeneral
TypeCO2 Incubator
Temperature RangeAmbient +5°C to 50°C
Temperature Control±0.1°C
CO2 Range0 to 20%
Humidity RangeUp to 95% RH
Interior Volume170 liters
Capacity6.0 cubic feet
CO2 Control Accuracy±0.1%
Power Supply50/60Hz
External Dimensions (WxDxH)620 x 715 x 905 mm
Internal Dimensions (WxDxH)490 x 505 x 690 mm (19.3 x 19.9 x 27.2 in)

Related product manuals