24
14. When the power is supplied (without battery)
Compressor H: Compressor H turns on with 2minutes (initial value) of delay after the
power was supplied. Compressor H runs with regardless of PV.
Compressor L:
In PV䋾SV (when the cascade sensor temperature is -34.1㷄 or lower),
compressor L turns on with 2minutes (initial value) of delay after the
power was supplied.
In PV䋾SV (when the cascade sensor temperature is -34.0㷄 or higher),
compressor L turns on with 2minutes (initial value) of delay after the
power was supplied and cascade temperature is reached at -34.1㷄.
*Note) Delay time of both compressor H and L cannot be set individually.
Setting data: The setting data initialized in F17 is retrieved in non-volatile memory.
15. Other specifications
(1) Lamp specifications:
<Control PCB>
DP1: Orange lamp
Goes off: High/low temp. alarm, sensor failure, power failure
Lit : Not in alert condition
DP2: Green lamp
Goes off: Compressor L turns off. (normal condition)
Lit : Compressor L turns on.
DP3: Red lamp
Goes off: Capillary heater turns off. (normal condition)
Lit : Capillary heater turns on.
DP4: Yellow lamp
Goes off: Inoperative
Lit: Inoperative
DP6: Green lamp
Goes off: Compressor H turns off. (normal condition)
Lit : Compressor H turns on.
<Display PCB>
DP51: Red lamp
Goes off: Not in alarm condition (normal condition)
Blinks : High/low temp. alarm (without delay), or sensor failure,
or power failure
DP52: Orange lamp
Goes off: Inoperative
Lit : Inoperative
Blinks : In F11 performed
DP53: Orange lamp
Goes off: Not in filter alarm (normal condition)
Lit : In filter alarm
Blinks : In F11 performed
DP54: Red lamp
Goes off: Inoperative
Lit : Inoperative
Blinks : In F11 performed
DP55: Orange lamp
Goes off: Length of battery used is 2.8yrs or less (normal condition)
Fan motor accumulation time is less than 5.6yrs
Lit : Length of battery used is 2.8yrs or over
Blinks : In F11 performed
Fan motor accumulation time is equal or more than 5.6yrs
DP56: Green lamp
Goes off: Inoperative