(3) Functions
of
test
mode
1.
Minute temperature display
When displaying the current temperature, the
value after the decimal point is also displayed.
2. Expansion
of
temperature setting range
Temperature can be set from
20"C to 150"C.
Error is prohibited by the sterilization limiter.
1
Displays
112.3"C.
Set value can be changed even after heating process.
3. Remaining time
display in exhaust process
Exhaust remaining time is
displayed on the Digital Display
II.
(in min.)
4. Pressure/Saturated steam pressure
display
1-
, L
I
By pressing the "TIMER" button, the current pressure is the Digital Display
I.
(in kPa)
5. Forced exhaust
12°31
I
By pressing the "A" and 'STOP" buttons at the same time during stand-by, the exhaust valve opens.
The forced exhaust can be
cancelled by pressing the "STOP" button.
6.
TImer 60-time speed operation
While the "TIMER" and the "SET/CONFIRM" button at the same time, the reservation timer, process
timer and
sterilization timer operate in 60-time speed. (It is preferable to perfonn the exhaust fast.)
7.
sterilization process starUcompletion signal output
Buzzer sounds at the start and
completion
of
the sterilization process.
(4)Functions
of
demonstration mode
1.
Power is not supplied to the heater.
2. Process timer and
sterilization timer operate
in
60-time speed.
3. When the power is
supplied to the heater, the temperature on the Digital Display rises automatically.
4. When the power is not
supplied to the heater, the temperature
on
the Digital Display lowers automatically.
(For safety sake, however, the temperature does not
lower than the actual temperature
of
TH1.)
5. Pressure
Abnonnal is not detected.
-
27-