Example: Enter 100 g as the upper limit of Class 1. Then enter 15%. When working 
with 3 classes, this yields the following weight classes:
  Class 0: up to the minimum load
   Class 1: > minimum load - 100 g
  Class 2: > 100 g – 115 g
  Class 3: > 115 g - maximum load
  When working with 5 classes, this yields the following weight classes:
  Class 0: up to the minimum load
  Class 1: > minimum load - 100 g
  Class 2: > 100 g – 115 g
  Class 3: > 115 g – 130 g
  Class 4: > 130 g – 145 g
  Class 5: > 145 g - maximum load
The values entered remain valid until deleted by pressing the c key or until 
overwritten by a new value. They remain saved after the scale is switch-off. 
  Preparation  t  Open the APPLIC. / APPLIC.2 / CLASS. menu.
  Available parameter settings   * = Factory setting
MIN.INIT Minimum load for initialization   3.6
  1 digit  1 digit  3.6.1*
  2 digit  2 digits  3.6.2
  5 digit  5 digits  3.6.3
  10 digT.  10 digits  3.6.4
  20 digT.  20 digits  3.6.5
  50 digT.  50 digits  3.6.6
  100 diG.  100 digits  3.6.7
  200 diG.  200 digits  3.6.8
  500 diG.  500 digits  3.6.9
  1000 dI.  1000 digits  3.6.10
CTRL.SET Activate SET control output   4.3.
  OUTPUT  SET output  4.3.1*
  OP.READY  Ready to operate  4.3.2
OUTP.ACT  Activation of outputs   4.7
  OFF  Off  4.7.1*
  ALWAY.ON  Always on  4.7.2
  STABIL.  On at stability  4.7.3
QTY. Number of classes    4.8.
  3  CLASS  3 classes  4.8.1*
  5  CLASS  5 classes  4.8.2
INPUT Parameter input    4.9. 
  WEIGHTS  Weight values  4.9.1*
  PERC.TAG  Percentage values  4.9.2
PRINT Automatic Printing  4.10. 
  manual  Off  4.10.1*
  automat  On  4.10.2
  )   t   To save the setting, press the ) key.
  ( ( ...  t   To exit setup: Press the ( key several times.
  Operating Instructions Combics Indicators  81
  Operation