Data Interface
Purpose
Your balance/scale is comes equipped
with an interface port for connection to a
computer or other peripheral device.
You can use an on-line computer to
change, start and/or monitor the
functions of the balance/scale and the
application programs.
Features
Type of interface: Serial interface
Operating mode: Full duplex
Standard: RS-232
Transmission rates:
150; 300; 600; 1,200; 2,400; 4,800;
9,600; 19,200 baud
Parity: Mark, space, odd, even
Character format:
1 start bit, 7-bit ASCII, parity,
1 or 2 stop bits
Handshake:
2-wire interface:
via software (XON/XOFF);
4-wire interface: via hardware handshake
lines (CTS/DTR)
Operating mode: SBI
Data output format of the balance/scale:
16 or 22 characters
54
Factory settings:
Transmission rate: 1,200 baud (5 1 4)
Parity: Odd (5 2 3)
Stop bits: 1 stop bit (5 3 1)
Handshake: Hardware, 2 characters
after CTS (5 4 2)
Operating mode: Standard SBI (5 5 1)
Print manually/automatically:
Manual at stability (6 1 2)