Competence Series/CP Series
Model CP423S-0CE CP323S-0CE CP323P-0CE CP153-0CE
Type BD BL 100 BD BL 100 BD BL 100 BD BL 100
Accuracy class
1
) K KKK
Maximum capacity, Max.
1
) g 420 320 80/160/320 150
Scale interval, d
1
) g 0.001 0.001 0.001/0.002/ 0.001
0.005
Tare range (subtractive) g < 100% of the maximum capacity
Verification scale interval, e
1
) g 0.01 0.01 0.01 0.01
Minimum capacity, Min.
1
) g 0.02 0.02 0.02 0.02
Response time (average) s < 1.5
Range of use
according to CD
1
) g 0.02–420 0.01–320 0.02–320 0.02–150
Allowable ambient
operating temperature °C +10…30 (50°F to 86°F)
Net weight, approx. kg 4.6
Pan size (inner diameter)* mm 110 d
Pan area* cm
2
120
Weighing chamber height
(from pan to sliding cover) mm 50
Dimensions (W+D+H) mm 213+342+153
AC power source/ V~ AC adapter STNG6 230 VAC or 115 VAC,
Power requirements +15%…–20% (protection rating IP20)
Frequency Hz 48 – 60
Power consumption (average) VA maximum: 16; average: 8
Approx. hours of operation
with the YRB05Z
rechargeable battery pack h 27
Selectable weight units Grams, carats
Built-in interface RS-232C-S/V24-V28; 7-bit;
parity: even, mark, odd, space;
transmission rates 150...19,200 baud;
1 or 2 stop bits; software/hardware handshake
1
) CD = Council Directive 90/384/EEC on non-automatic weighing instruments;
applicable to the European Economic Area