12
Connecting Electronic Peripheral Devices
§ Make absolutely sure to unplug the balance/
scale from AC power before you connect or
disconnect a peripheral device (printer or PC)
to or from the interface port.
Warmup Time
To deliver exact results, the balance/scale
must warm up for at least 30 minutes after
initial connection to AC power or after a
relatively long power outage.  Only after this
time will the balance/scale have reached the
required operating temperature.
Using Verified Balances/Scales as Legal
Measuring Instruments in the EU*
$ The balance/scale must warm up for at 
least 24 hours after initial connection to 
AC power. 
Leveling the Balance/Scale
Purpose
– To compensate for unevenness at the place
of installation.
– To achieve perfectly horizontal positioning
of the balance/scale for consistent
reproducibility
Always level the balance/scale again after
any time it has been moved.
* including the Signatories of the Agreement
on the European Economic Area