Calibration/Adjustment Sequence
You can configure the operating menu 
so that:
– adjustment automatically follows calibra-
tion in a single operation (1 10 1), or 
– the operator chooses whether to end 
the calibration/adjustment routine or
have adjustment performed (1 10 2) 
If there is no deviation, or if the
difference is within the effective require-
ments for accuracy of measurements, 
it is not necessary to adjust the balance/
scale.  In this case, you can end the
calibration routine following calibration.
Two keys are active at that point:
– q = start adjustment
– c = end the calibration/
adjustment routine
Factory Settings
Calibration adjustment mode for 
models without built-in motorized
calibration weight: 
External calibration (1 9 1)
Calibration adjustment mode for models
with built-in motorized calibration weight: 
Internal calibration (1 9 3)
Calibration/adjustment sequence: 
Adjustment automatically follows calibration 
in a single operation (1 10 1)
Weight unit for calibration: grams (1 11 1)
ISO/GLP-compliant printout: off (8 10 1)