Weighing in Percent
Menu code: 2 1 5
Display symbol: %
Purpose
This application program allows you to
obtain weight readouts in percent which
are in proportion to a reference weight.
Features
The minimum load is equal to one digit,
defined according to the resolution of the
active weight unit.
Press and hold the v key (2 seconds)
to set the reference percentage
Storage parameter (rounding-off factor)
for storing the reference weight
to calculate the percentage can be
configured.
Configuration of decimal places displayed
with a percentage.
Optional automatic output of the
reference weight “Wxx%” and reference
percentage to the data interface port
when the menu code 7 1 2 (print
application parameters) is set.
42
Long-term storage of the last reference
percentage “pRef” entered.
Toggling between percentage and weight
by pressing v.
Function Keys
v:
Begin calculation of percentage
> Current weight value stored as reference
weight “Wxx%” to be loaded at initialization.
c:
End application program;
clear initialization data
Changing the reference percentage:
§ Press and hold v (2 sec.)
> Current reference percentage is displayed.
$ Press v briefly to change the value;
press repeatedly until the desired reference
percentage is displayed.
Quantities to choose from: 1, 2, 5, 10, 20,
50, 100.
§ Store setting in long-term memory:
Press and hold v.