EasyManua.ls Logo

Sartorius ED2202S - Application Programs: Weighing in Percent; Weighing in Percent Functionality

Sartorius ED2202S
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Weighing in Percent
Display symbol: %
Purpose
This application program allows you to
obtain weight readouts in percent which
are in proportion to a reference weight.
Changing the Reference Percentage
Activate function:
Press the b key
Select the desired reference (1 to 100):
In increments of 1: Press the b key
briefly
In increments of 10: Press and hold the
b key.
The percentage is stored in battery-backed
memory.
Preparation
§ Select the Weighing in percent
application in the menu:
see “Configuration.”
§ Set the following parameters:
Applic. Application program
Percent Weighing in percent
Dec.Plcs. Decimal places
None No decimal places
o 1 Dec. Pl. 1 decimal place
2 Dec. Pl. 2 decimal places
3 Dec. Pl. 3 decimal places
o = Factory setting
Printout: Weighing in percent
pRef 100 : Reference
percentage
Wxx% 111.6 g : Reference weight
net xx% for selected
reference percentage
Prc + 94.9 pcs : Calculated reference
percentage
28

Related product manuals