ISO/GLP-compliant Printout/Record
Features
You can have device information, ID
texts and date and time printed before
(GLP header) and after (GLP footer) the
values of a weighing series. These
parameters include: 
GLP header:
– Date
– Time at beginning of measurement
– Balance/scale manufacturer
– Balance/scale model
– Balance/scale serial number
– Software version number
– Identification number of the current
sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
!Operating the Balance/Scale 
with a Verifiable ISO/GLP Printer:
§ Connect a Sartorius data printer
designed for ISO/GLP documentation
(e.g., the YDP03-0CE printer) to the 
balance/scale.
Configuration
§ Setting menu codes for the printout
(see “Configuration”):
– ISO/GLP-compliant printout or record
only for calibration/adjustment: 
Setup: Prnt.Out: GLP: Cal.-Adj.
(menu code 1. 6. 7. 2) or 
ISO/GLP-compliant printout or record
always on: Setup: Prnt.Out: GLP:
Always on (code 1. 6. 7. 3)
– Line format for printout: include data
ID codes (22 characters; 
factory setting):  
Setup: Prnt.Out: Format: 22 Char.
(menu code 1. 6. 6. 2) 
– Formats for time:
setup: prnt.0ut: time: 24 h
(menu code 1. 6. 8. 1) or
setup: prnt.oUt: time: 12 h
(menu code 1. 6. 8. 2) 
– Formats for date:
setup: prnt.0ut: date: dd.mmm.yy
(menu code 1. 6. 9. 1) or
setup: prnt.oUt: date: mmm.dd.yy
(menu code 1. 6. 9. 2) 
!No ISO/GLP-compliant record is output
if any of the following settings are 
configured:
Setup: Prnt.Out  Print: Aut. W/o or
Aut.With (menu code 1. 6. 1. 3, 1. 6. 1.
4, ) or Format: 16 Char. (menu code 
1. 6. 6. 1) 
Function Keys
Transfer header and first measured
value: press r
> The header is included with the first
printout/data record. 
To output header and reference data
automatically when an application 
program is active: press u
Exit the application: 
1) To send the GLP footer: 
press c
2) Quit application program: 
press c again
44