Below-Balance/Scale Weighing
A port for a below-balance/scale weighing hanger
is located on the bottom of the balance/scale.
$ Below-balance/scale weighing is not permitted in
legal metrology.
§ Open cover plate on the bottom of the
balance/scale.
Important: set the balance/scale on its side to access
the cover plate. DO NOT turn the balance/scale
upside-down!
§ Using the built-in hook 1: Attach the sample
(e.g., using a suspension wire) to the hanger.
$ Install a shield for protection against drafts
if necessary.