Internal Calibration/Adjustment
(Only on Models with a Built-in Motorized Calibration Weight)
Standard feature on the following models:
Models with a readability of 0.1 mg
ED…-CW models
GK… models
GW… models
Verified models (with the …CE suffix)
Set the following parameters:
Setup: Bal.Scal.: Cal.-Adj.: Cal.Int. (menu code 1.1. 9. 4)
The built-in motorized calibration weight is applied and removed automatically for inter-
nal calibration.
Step Key (or instruction) Display
1. Tare the balance/scale
w 0.0 g
2. Start calibration
Q Cal.Int.
The built-in weight is
Cal. run.
applied automatically
3. Calibration/adjustment executed
Cal.end
4. The built-in weight is removed 0.0 g
16