EasyManua.ls Logo

Sartorius ME235P - SQmin Function; Factory-set parameters

Sartorius ME235P
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
115
Example
Create a new product base data record for initializing the checkweighing program, including: target value, minimum, maximum
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Extra function (F4): Product data memory
Setup: Application parameters: Application 2: Checkweighing
Step Press key(s) (or follow instructions) Display/Output
1. If necessary, turn on the balance; e
then enter the settings given above
2. In the Checkweighing application,
Param. soft key
toggle to the input mode for target,
minimum and maximum values
3. Enter target: 170 g; see the Example
minimum: 165 g; for Checkweighing,
maximum: 180 g steps 5 through 9
4. Toggle to display of product data
ProDat soft key
(existing data records are displayed;
in this example, 3 data records have
been stored)
5. Enter a name for the new data a
record (here: CHW01)
ABCDEF soft key, C soft key
GHIJKL soft key, H soft key
STUVWX soft key, W soft key
0 1
6. Store current Checkweigh
New soft key
parameters as a data record
7. Confirm storage function
Store soft key
8. Exit product data display
oo soft key
oo q l
CHECKWEIGH: 0.00000 g
Target: Setp= + 0 g
Minimum: Min = + 0 g
Maximum: Max = + 0 g
oo Q l
CHECKWEIGH: + 170.00000 g
Target: Setp= +170.00000 g
Minimum: Min = +165.00000 g
Maximum: Max = +180.00000 g
oo Lösch o Q Store
PROD.DATA: NEW: KW01
CHECKWEIGH
Setp= +
Min = +
Max = +
Lim-=
Lim+=
170.00000 g
165.00000 g
180.00000 g
3 %
6 %
g
Max 210 g
0%
CHCKW.: n = 1 Setp= +170.00000 g
d=0.01mg
100%
16948765
F
S
a
s
Cal ShowProDatParam. Net
D
H
S
T
oo Lösch o Q New
PROD.DATA:
Verrechnen8
Zählen13
PPOZENTWÄG.
Wxx%
pRef
68.75 g
100 %
CHW01
oo Delete Load q Q Change
PROD.DATA: PERCENT WGH
PERCENT WGH40
CALCULATION8
COUNTING13
Wxx%
pRef
168.75432 g
100 %
oo Delete Load q Q Change
PROD.DATA: Data stored
CHW01
PERCENT WGH40
CALCULATION8
COUNTING13
Setp= +
Min = +
Max = +
Lim-=
Lim+=
170.00000 g
165.00000 g
180.00000 g
3 %
6 %

Table of Contents

Related product manuals