EasyManua.ls Logo

Sartorius ME235P - Page 135

Sartorius ME235P
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Format (Line Format)
You can output the values displayed in the
line for measured values and the weight
unit with or without a data ID code
Example: Without data ID code
+ 253 pcs
Example: With data ID code
Qnt + 253 pcs
Configure this parameter in the Setup
menu (Setup: Printout: Line format).
The output with a data ID code has
16 characters; without a data ID code,
22 characters.
Output Format With 16 Characters
Display segments that are not activated
are output as spaces. Characters without
a decimal point are output without a
decimal point.
134
The following characters can be output, depending on the characters displayed on the balance:
Normal Operation
Position 1 2345678910111213141516
+DDDDDDDDD* UUUCRLF
or ......... ***
or **********
*: Space
D: Digit or letter
U: Unit symbol
CR: Carriage return
LF: Line feed
Special Codes
Position 1 2345678910111213141516
************CRLF
or ******A******* CRLF
or ******AB ******CRLF
and only upon request with ESC w0 (no print command):
******S*XXXYYYCRLF
or ******I*XXX***CRLF
*: Space
AB = – –: Final readout mode A = C: Cal/adjustment
A = H: Overload S: Draft shield status
AB = H H: Overload in checkweighing I: Ionizer
A = L: Underload Y,Y,Y = Draft shield doors
AB = L L: Underload in checkweighing
XXX =
Decimal value calculated from individual status bits:
Decimal Binary Control
value value information
1
Bit0 = 0: No error/ionizer off
Bit0 = 1: Draft shield error/ionizer on
2
Bit1 = 0: Draft shield motor off
Bit1 = 1: Draft shield doors in motion
8
Bit3 = 0: Learning function off
Bit3 = 1: Learning function on
16
Bit4 = 0: At least one draft shield door open
Bit4 = 1: All draft shield doors closed
64
Bit6 = 0: Motorized draft shield operation
Bit6 = 1: Manual draft shield operation
Example for ME215/235/254/414/415/614:
R, M, L = COO: right door closed, middle and left doors open
R, M, L = OCC: right door open, middle and left doors closed
Example for ME5, SE2:
W 008210
Angle of aperature:
The draft has turned itself by 210°.
Control information:
0 Bit 6 – Motorized draft shield operation
+0 Bit 4 – Draft shield open
+8 Bit 3 – Learning function on
+0 Bit 1 – Draft shield motor off
+0 Bit 0 – No error
=08

Table of Contents