Preparation
● Select the balance function for “calibration/adjustment”:  press s
● To select the 
Balance/scale functions: press the O soft key
● Select
Calibration/adjustment:  press the O soft key
Calibration/ CAL key function Ext. cal./adj.: default wt.
adjustment Ext. cal./adj.: user-defined wt.
Internal cal./adjustment
Internal linearization
Key blocked
reproTEST
ο Selection mode
Cal/adjustment  ο Calibrate, then auto adjust
sequence Calibrate, then manual adjust
isoCAL function Off
Only adjustment prompt
On and reset application
isoCAL and linearization on*
ο On without resetting app.
Start automatic  User-def. Adj. time1
adjustment adjustment times Adj. time2
Adj. time3
ο isoCAL
Print GLP/GMP ο Automatic if GLP is selected
adjustment record On request, from record memory
Parameter for Weight ID (Wt. ID)
external weight Cal./adj. wt.:
ο = factory setting
* this deletes (resets) the application
● Save settings and exit Setup menu: press the 
oo soft key
Releasing Access to 
External Calibration
● Remove the cover plate from the back
of the balance housing
● Move switch 1 in the direction 
of the arrow
> Switch left: 
external calibration accessible 
Switch right: 
external calibration blocked
> Note: 
Do not move switch 2
48