Example
Determining the diameter of metal wires and filaments (such as a filament used in a light bulb) 
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Diameter determination
Setup: Application parameters: Extra functions (F5): Product data memory
Step Press key(s) (or follow instructions) Display/Output
1. Turn on the balance, if not on e
2. Toggle to the parameter
Param. soft key 
display
3. Enter name for COMPANY a ...
(in this example, Sartorius) See also page 42
4. Confirm input a, 
l soft key
5. Enter density of filament
q soft key, 
(in this example, tungsten) 1 9 . 2 5, 
l soft key
6. Enter filament length in millimeters, then  2 0, 
l soft key, 
exit parameter settings
oo soft key
7. Start diameter determination
Start soft key
The printout mode can be switched off in 
SARTORIUS
the Setup menu (Setup: Printout: Application- RhoW 19.25 g/
defined output: Auto print upon l20mm
initialization: Off) # dec. 3
95