119
Example
Perform a weighing procedure with the “DKD uncertainty of measurement” application
Settings:
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Display: PA (process accuracy)
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Print: PA (process accuracy)
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Display: Input: 3.00000 (factory setting)
Step Press key(s) (or follow instructions) Display/Output
1. Turn on the balance, if not on, and e
configure the settings as indicated above
2. Place a container for a sample on the t
balance and tare
3. Measure weight of sample Add sample to container
4. Print weight p/r U 0.000292 g
PA 0.00087 %
N +100.54292 g
5. Weigh next sample (if any)