Preparation
Example: Set the parameters for calibration and adjustment; e.g., with manual calibration/adjustment, isoCAL off
Step Press key(s) (or follow instructions) Display/Output
1. Switch on the balance, e Sartorius logo and self-test
if not already on
2. Select the Setup menu s
3. Select “Balance/scale functions”
O soft key
4. Select “Calibration/adjustment”
O soft key
5. Select CAL key function
O soft key
o = last setting selected
6. Select desired function and
Q soft key, repeatedly,
confirm (e.g., “Internal cal./adj.”) if necessary l soft key
7. Exit CAL key function
o soft key
8. Select “Cal./adjustment sequence”
q soft key
49
g
Max 210 g
0%
d=0.01mg
100%
000000
S
Cal