11
Setting Up the ME5 or SE2 Balance
§ Place the components below on the weigh cell base in the order given:
Shield disk
Weighing pan
Please note: after placing the weighing pan on the base, turn the pan slightly towards the
left then the right while pressing down on it to lock.
Interior draft shield (only for SE2)
Draft shield: center the hole over the pan (see arrows)
Setting Up the ME5-F or SE2-F Balance
§ Place the components below on the weigh cell base in the order given:
Shield disk
Interior draft shield ring
50 mm dia. filter pan or weighing pan (or, optionally, the 75 or 90 mm dia. filter pan)
Please note: after placing the pan on the base, turn the pan slightly towards the left then
the right while pressing down on it to lock
Interior draft shield (only for SE2)
Draft shield cover
Setting Up the Filter Balance for Left-handed Persons:
§ Remove the draft shield cover
§ Detach the pin on the right and re-attach on the left
ME5, SE2:
ME5-F, SE2-F: