151
Verified Models with EC Type Approval
Model ME415S-0CE ME235S-0CE ME215S-0CE ME235P-0CE ME215P-0CE
Type BE BK BE BK BE BK BE BK BE BK
Accuracy class* kkkk k
Scale interval d* mg 0.01 0.01 0.01 0.01/0.02/0.05 0.01/0.02/0.05
Max. weighing capacity* g 410 230 210 60/110/230 60/110/210
Verifcation scale interval e* mg 1 1 1 1 1
Min. capacity* mg 1 1 1 1 1
Tare range (subtractive) <100% of the max. weighing capacity
Application range* g 0.001-410 0.001-230 0.001-210 0.001-230 0.001-210
Response time (average) s < 8 < 8 < 8 < 8 < 8
Allowable operating temperature range
1
) 273 … 313 K (0 … +40°C, 32°F … 104°F with “isoCAL” function
External calibration weight g 2+ 200 (E2) 200 (E2) 200 (E2) 200 (E2) 200 (E2)
(of at least accuracy class...)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update rate
(depends on filter level) 0.2 – 0.4
Pan size mm d 90
Balance dimensions (W + D + H) mm 252 + 533 + 292
Weighing chamber height mm 239
Net weight, approx. kg 11.1
Dust and water protection rating of the
balance housing according to EN 60529 IP32
Power connection Using wide-range AC adapter for voltage ratings of 100 V to 240 V
Nominal frequency 50 60 Hz
Power consumption max. 35 VA
Hours of operation with fully charged
YRB05Z external battery pack, approx. 10 h
DC nominal supply voltage 10.5 to 25 Vdc
Selectable weight units Grams, carats and milligrams
Selectable application programs Toggle weight units, counting, weighing in percent, animal weighing,
recalculation, calculation, density determination, differential weighing,
air buoyancy correction, air density determination, checkweighing,
time-controlled functions, totalizing, formulation, statistics,
2nd tare memory, identification codes, product data memory
1
) = With the isoCAL function deactivated, the verified balance can be used only within the limited temperature range
(can be modified only by the Sartorius Service Center): For balances of accuracy class k: +15°C to +25°C (59 to 77°F)
* = according to Council Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area