16
2. Press Mode until your digital display indicates
mV mode.
3. Press Standardize to enter an mV standard and
read relative mV.
4. When the signal becomes stable, or when you
press Enter, the current absolute mV value (offset)
becomes zero relative millivolts.
[Mode]
[Standardize]