Practical Example: Totalizing Symbol displayed: Σ Code 2 7 2*
and “Counting” Symbol displayed: P Code 2 1 4*
Totalizing S/a key
Read Totalizing Memory s/r key
User guide display when you press i/i key
+ s/r key: Sn, St, Sb, SC
Application: Totalizing bulk parts of equal weight or weight values, whereby
the total value displayed can exceed the capacity of the scale many
times over.
Note:
In this example, the containers all have the same weight.
Special menu code settings used in this example:
Function Code
Automatic output of totalizing
application data: on 3 13 1
Output all parameters 7 1 2
Step/Key Display Data Output
Place the empty container +
4.2 gcs
Press the =/z key
0.0 gcs
Press the g/W key,
Select reference sample
quantity, e.g. 20 pcs
0.0 gcs
Place 20 pieces
in the container +
17.2 gcs
Press the u/w key,
begin counting +
20 pcs
Data output nRef + 20 pcs
wRef + 0.8700 g
Press the S/a key +
20 pcs
nRef + 20 pcs
to store value in the wRef + 0.8700 g
totalizing memory. n1
Transaction counter: 1 N + 17.7 g
1st partial value Qnt + 20 pcs
* = factory setting