Satelec Cocoon Spray - 2.6 Troubleshooting

Satelec Cocoon Spray
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
FAULT
No operation
No spray
No water at tip but
vibration present
Insufficient power
Low vibration
No ultrasonic output
Water leak between
handpiece and cord or at
front/back junction of
handpiece
REMEDY
Check mains outlet
Return to customer
service
Dry any moisture, especially
on electrical contacts
Check water supply
Check pressure
Clean or replace filter
Return to customer
service
Remove obstruction
Adjust spray
Replace tip
See tip utilization method
Wipe electrical contacts
dry
Tighten with wrench
Clean cord contacts
Return to customer
service for cord
replacement
Replace seal
POSSIBLE CAUSE
Wrong mains cord connection
Blown fuse
Presence of fluid or moisture between
handpiece and cord
Wrong water tube connection
No water pressure
Clogged filter or defective solenoid
valve
Blocked tip
Incorrect spray adjustment
Worn or deformed tip
Incorrect use: wrong angle of
application or inadequate pressure on
tooth
Presence of fluid or moisture between
handpiece and cord
Incorrect tip tightening
Defective connector contact
Cut wire in handpiece cord
Worn sealing gasket (1.15 x 1) in
handpiece