Satelec Cocoon Spray - 2.6 Troubleshooting

Satelec Cocoon Spray
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
FAULT
No operation
No spray
No water at tip but
vibration present
Insufficient power
Low vibration
No ultrasonic output
Water leak between
handpiece and cord or at
front/back junction of
handpiece
REMEDY
• Check mains outlet
• Return to customer
service
• Dry any moisture, especially
on electrical contacts
• Check water supply
• Check pressure
• Clean or replace filter
• Return to customer
service
• Remove obstruction
• Adjust spray
• Replace tip
• See tip utilization method
• Wipe electrical contacts
dry
• Tighten with wrench
• Clean cord contacts
• Return to customer
service for cord
replacement
• Replace seal
POSSIBLE CAUSE
• Wrong mains cord connection
• Blown fuse
• Presence of fluid or moisture between
handpiece and cord
• Wrong water tube connection
• No water pressure
• Clogged filter or defective solenoid
valve
• Blocked tip
• Incorrect spray adjustment
• Worn or deformed tip
• Incorrect use: wrong angle of
application or inadequate pressure on
tooth
• Presence of fluid or moisture between
handpiece and cord
• Incorrect tip tightening
• Defective connector contact
• Cut wire in handpiece cord
• Worn sealing gasket (1.15 x 1) in
handpiece