12
FAULT
Motor will not run after the
pedal has been activated
No water output at contra-
angle
The luminous switch does
not light up
Rotation direction does not
match selection
REMEDY
Plug in pedal cable jack
correctly
Turn it to desined position
Plug in motor cable jack
correctly
Tighten motor cable jack
Slightly raise the 2 blades
by means of a screwdriver
Return to SATELEC
Customer Service
Change cable
Change pedal
Change motor
Check water input
Check pressure
Clean or change filter
Return to SATELEC
Customer Service
Remove obstruction
Remove obstruction
Replace fuses
Check micromotor for
correct assembly onto
connecting cable (red
mark)
POSSIBLE CAUSE
Wrong pedal cable connection
Motor rotation reversal button in
central position
Wrong motor cable connection
Motor cable jack loose on cable
connector
Cable connector contact blades too
flattened
Control box malfunction
Cut cable
Defective pedal
Motor failure
Water tube incorrectly connected
Insufficient water pressure
Blocked filter of faulty electrovanne
Contra-angle blocked
Micromotor water outlet connector
blocked
Fuses blown
Incorrect micromotor connection