12
FAULT
Motor will not run after the
pedal has been activated
No water output at contra-
angle
The luminous switch does
not light up
Rotation direction does not
match selection
REMEDY
• Plug in pedal cable jack
correctly
• Turn it to desined position
• Plug in motor cable jack
correctly
• Tighten motor cable jack
• Slightly raise the 2 blades
by means of a screwdriver
• Return to SATELEC
Customer Service
• Change cable
• Change pedal
• Change motor
• Check water input
• Check pressure
• Clean or change filter
• Return to SATELEC
Customer Service
• Remove obstruction
• Remove obstruction
• Replace fuses
• Check micromotor for
correct assembly onto
connecting cable (red
mark)
POSSIBLE CAUSE
• Wrong pedal cable connection
• Motor rotation reversal button in
central position
• Wrong motor cable connection
• Motor cable jack loose on cable
connector
• Cable connector contact blades too
flattened
• Control box malfunction
• Cut cable
• Defective pedal
• Motor failure
• Water tube incorrectly connected
• Insufficient water pressure
• Blocked filter of faulty electrovanne
• Contra-angle blocked
• Micromotor water outlet connector
blocked
• Fuses blown
• Incorrect micromotor connection