Program 10: Gate control dynamic 1
The intensity continuously changes and the two channels operate alternately
(massage eect).
Frequency: 80 Hz
Pulse width: 150 µs
Intensity modulation: 1 s rise, 1 s fall
Therapy time: 30 min (10 to 90 min)
Indications: Especially for acute back
pain, lower back pain, sciatic pain
This program is recommended for the
treatment of back pain. The massage
eect provides a comfortable and pleasant
stimulation. The so called Gate Control reflex
which is achieved through this kind of stimulation is a natural reflex that you achieve e.g. by
rubbing a painful area with your hands to calm acute pain. To achieve the massage eect it
is essential to use both channels of the TENStem eco basic.
Program 11: Gate control dynamic 2
The intensity continuously changes two channels operate alternately
(massage eect). See illustration program 10.
Frequency: 80 Hz
Pulse width: 150 µs
Intensity modulation: 0.25 s rise, 0.25 s fall
Therapy time: 30 min (10 to 90 min)
Indications: Heavy legs (venous return)
Program 11 is similar to program 10, but the movement of the massage waves is
faster. It is used to improve the venous return for patients who suer from heavy
legs. To achieve the massage eect it is essential to use both channels of the TENS-
tem eco basic.