Program 13: Gate control
Both channels work with the same frequency and pulse width.
Frequency: 1 - 120 Hz
Pulse width: 75 - 300 µs
Therapy time: 10 - 90 min
Program 14: Muscle stimulation
The intensity continuously changes according to the indicated parameters.
Frequency: 1 - 120 Hz
Pulse width: 75 - 300 µs
Rising ramp: 2 s
Working phase: 5 s
Falling ramp: 1 s
Pause: 12 s
Therapy time: 10 - 90 min
Program 15: Gate control dynamic
The intensity continuously changes and the two channels operate alternately
(massage eect).
Frequency: 1 - 120 Hz
Pulse width: 75 - 300 µs
Intensity modulation: 1 s rise, 1 s fall
Therapy time: 10 - 90 min
Program 16: Iontophoresis, Nerv block (from series E)
Both channels work with the same frequency and pulse width. The current is
monophasic.
We do not recommend to use P16 for more than 20 minutes. Check the skin
underneath the electrodes after the stimulation.
Frequency: 1 - 120 Hz
Pulse width: 75 - 300 µs
Therapy time: 20 min




























  










 






 


















Intensity
Time




























  










 






 


















working time
Pause
rising time
falling time
Intensity




























  










 






 


















Intensity