inlet
valves
drain
hose
hot water inlet hose
cold water
inlet hose
Hydrim
®
C51w Installation Instructions
order number 01-108391S
To ols and supplies required
to install the Hydrim:
•Slot screwdriver • Channellocks
Utility Hook-ups
Proposed Installation Configurations:
If the Hydrim is installed in a Sterilization Center, the
manufacturer of the Sterilization Center should allow a 3"
(75 mm) space at the top and sides of the Hydrim. This will
facilitate installation, leveling, and service access to
the Hydrim.
Do not move the Hydrim into place by maneuvering the
open wash chamber door. This may cause to door to
become misaligned and leak.
* unit will function with water pressure down to 0.5 bar / 7 psi.
Ensure that Hydrim Cleaning Solution (instrument wash
chemical) is available. All other supplies are included
with the Hydrim unit.
Specifications
North America Europe
Voltage: 208 - 240 V 220 - 240 V
Frequency: 60 Hz 50 Hz
Rated load: 2 kW 2 kW
Circuit breaker: 10 A per phase 10 A per phase
Dimensions / Mass:
Height: 475 mm / 18.75"
Width: 600 mm / 23.75"
Depth: 460 mm / 18.25"
Depth with
door open: 780 mm / 30.75"
Weight: 34 kg / 75 lbs
Running Noise: 60 dBA
Hose / Length /Max. Distance Water Shut
Cord Diameter Distance from Pressure -off
inlet / drain (optimal)* valve
Hot inlet 1.9 m / 6.2 ft. 1.5 m / 5 ft. 1-10 bar / Yes
2 cm / 3/4" 14.5-145 psi
Cold inlet 1.9 m / 6.2 ft. 1.5 m / 5 ft. 1-10 bar / Yes
2 cm / 3/4" 14.5-145 psi
Drain 1.5 m / 5 ft. — — —
2 cm / 3/4"
Electrical 1.8 m / 6 ft. — — —
AWG 18-3
Installation Option 1. Cabinet / Steri Center
Sliding Shelf
Installation Option 2. Sliding Shelf
Installation Option 3. Counter Top