EasyManuals Logo

SciCan hydrim c51w Installation Instructions & Owner's Manual

SciCan hydrim c51w
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
inlet
valves
drain
hose
hot water inlet hose
cold water
inlet hose
Hydrim
®
C51w Installation Instructions
order number 01-108391S
To ols and supplies required
to install the Hydrim:
•Slot screwdriver Channellocks
Utility Hook-ups
Proposed Installation Configurations:
If the Hydrim is installed in a Sterilization Center, the
manufacturer of the Sterilization Center should allow a 3"
(75 mm) space at the top and sides of the Hydrim. This will
facilitate installation, leveling, and service access to
the Hydrim.
Do not move the Hydrim into place by maneuvering the
open wash chamber door. This may cause to door to
become misaligned and leak.
* unit will function with water pressure down to 0.5 bar / 7 psi.
Ensure that Hydrim Cleaning Solution (instrument wash
chemical) is available. All other supplies are included
with the Hydrim unit.
Specifications
North America Europe
Voltage: 208 - 240 V 220 - 240 V
Frequency: 60 Hz 50 Hz
Rated load: 2 kW 2 kW
Circuit breaker: 10 A per phase 10 A per phase
Dimensions / Mass:
Height: 475 mm / 18.75"
Width: 600 mm / 23.75"
Depth: 460 mm / 18.25"
Depth with
door open: 780 mm / 30.75"
Weight: 34 kg / 75 lbs
Running Noise: 60 dBA
Hose / Length /Max. Distance Water Shut
Cord Diameter Distance from Pressure -off
inlet / drain (optimal)* valve
Hot inlet 1.9 m / 6.2 ft. 1.5 m / 5 ft. 1-10 bar / Yes
2 cm / 3/4" 14.5-145 psi
Cold inlet 1.9 m / 6.2 ft. 1.5 m / 5 ft. 1-10 bar / Yes
2 cm / 3/4" 14.5-145 psi
Drain 1.5 m / 5 ft.
2 cm / 3/4"
Electrical 1.8 m / 6 ft.
AWG 18-3
Installation Option 1. Cabinet / Steri Center
Sliding Shelf
Installation Option 2. Sliding Shelf
Installation Option 3. Counter Top

Other manuals for SciCan hydrim c51w

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SciCan hydrim c51w and is the answer not in the manual?

SciCan hydrim c51w Specifications

Operating Environment IconOperating Environment
maximum relative humidity80% for temp up to 31˚C
maximum relative humidity for higher temperature50% for temp up to 40˚C
operating temperature range-5˚C to 40˚C
maximum altitude2000m
Power IconPower
north america voltage208-240 V 60 Hz 10 A
europe voltage220-240 V 50 Hz 10 A
mains supply tolerance+ / -10% of nominal
Other Specifications IconOther Specifications
running noise60 dBA
hot water connection temperature50-70˚C
rinse aid dispenser capacity60 ml
water softener salt capacity0.5 kg
filling system safety maximum3.5 L
wash temperature50˚C +/- 5˚C
drying system heater1 kw
Weight and Dimensions IconWeight and Dimensions
height475 mm
width600 mm
depth460 mm
depth with door open780 mm
weight34 kg