Step 3: Setup Unit ID.
a) Power up unit while holding the Stop button pressed. The user
menu is going to be displayed.
b) By using Unwrapped and Wrapped keys scroll down to Unit ID
Setup. Press Rubber and Plastic key.
c) Setup Unit ID by pressing Unwrapped, Wrapped, and Rubber and
Plastic Keys. Press stop when done.
Step 4: Attach Statim USB adapter
a) Make sure that both Statim and Statim USB adapter are Off.
b) Connect Statim USB adapter to Statim unit by using the serial
cable.
c) Power up the USB Data Logger
d) Power up Statim unit.
e) LCD Display will show:
f) Insert USB memory
g) After a few seconds the LCD display changes to
HH:MM DD/MM/YYYY
“MSD NOT DETECTED”/”INSERT FLASH/MSD”/”SELECT A CYCLE”
HH:MM DD/MM/YYYY
USB/FLASH DETECTED/SAFELY REMOVE MSD/SELECT A CYCLE