EasyManua.ls Logo

SciCan STATIM 2000/S - Set Cycle Counter; Lock Cycles

SciCan STATIM 2000/S
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
To start the Chamber thermocouple Calibration cycle, first
press and release the Unwrapped key followed by pressing
the Start key.
Wrapped Sends the last printout to the printer. The printer must be
connected and selected for this feature to work.
Rubber and Plastics Start the Burn-in cycle
Stop Exit to normal mode of operation
Keypad functions when calibration cycle is running:
Unwrapped Increment current selected offset.
Wrapped Decrement current selected offset
Rubber and Plastics Select other offset
Stop Interrupt calibration cycle
2.2. Set cycle counter
Adjust the recorded number of cycles ran
Cycle #
000000
Keypad
Unwrapped Decrement current digit
Wrapped Increment current digit
Rubber and Plastics Move to next digit
Stop Save and exit to normal mode of operation
2.3. Lock cycles
Expands or contracts the selection of cycle parameters for each cycle button.
Keypad
Unwrapped Toggle to Locked/Unlocked mode
Stop Save and exit to normal mode of operation
Display after toggling to Unlocked mode
Cycles UNLOCKED
Available cycles
Display after toggling to Locked mode
Cycles LOCKED
Selected cycles
In the “Cycles LOCKED” mode, each cycle button will represent one set of cycle
parameters. In the “Cycles UNLOCKED” mode, each cycle button can represent
multiple sets of cycle parameters. After toggling to each mode, the corresponding