EasyManua.ls Logo

SciCan STATIM 2000/S - Page 44

SciCan STATIM 2000/S
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10. Statim 5000 Trouble-shooting – Cycle Faults
42
Cycle Fault
Number
Description of Fault
Suggested steps for Correction of fault
Cycle Fault #1
The cassette temperature failed to reach 95°C
within a time-out period.
This fault may be caused by a large cassette leak in
conjunction with an extremely large load or a blown
thermal fuse caused by weak pump delivery and/or
contaminated steam generator or a faulty steam
generator triac. If the triac has failed, the thermal fuse
may have also failed. A faulty steam generator (steam
generator resistance failed open) could also cause this
fault.
Cycle Fault #2
Software Revisions
R1xx or R2xx only
The cassette temperature failed to increase from
95°C to 100°C within a time-out period.
This may be cause by a major steam leak from the
cassette, incorrect chamber temperature calibration or
an extremely large load.
Cycle Fault #3
The cassette has failed to pressurize and achieve a
temperature of 110°C within a time-out period
This may be caused by a faulty cassette seal, a
damaged cassette or faulty solenoid valve (failed to
close). Verify check valve and pressure relief valves for
leaks. For PCB rev other than rev. 7.xx verify Vref to be
2.520V± 0.001V.
Cycle Fault #4
The cassette has failed to achieve sterilization
conditions within a timeout period of the chamber
first reaching 110°C.
This may be caused by a faulty cassette seal, a
damaged cassette, improperly installed copper tubing,
a failed or faulty solenoid valve (failed to close) , or
improper alignment of the probe bracket. Verify check
valve and pressure relief valves for leaks. For PCB rev
other than rev. 7.xx verify Vref to be 2.520V± 0.001V.
Cycle Fault #5
Software Revision
R1xx only
The software causes the pump to activate at
predetermined times. If a request to pump water
occurs outside of the predetermined time Cycle
Fault #5 occurs.
This may be caused by a leaky cassette, a kink in the
steam generator inlet tube, a leaking solenoid outlet
valve, a weak pump and a very large load in the
cassette.
Cycle Fault #6
Software Revisions
R2xx or R4xx
The software has detected the filtered steam
generator temperature 6°C greater than the
chamber during the sterilizing phase of a cycle
The most probable causes are an obstruction caused
by seal misalignment or foreign material in the seal
channel, a need for thermocouple or steam generator
calibration, an old or contaminated steam generator.