15
Default value is 0 or OFF.
Values available are:
0, 5, 10, 15 ... 55, 60 (minutes)
11. Pulse Recovery Time
This changes the time period allowed in making single shot energy measurements. Typically, the time it
takes to recover from a pulse is four to six times the calorimeters thermal time constant. The typical time
for each calorimeter size is given below.
Table 6. Calorimeter Pulse Recovery Time
AC25XXS
60 seconds
AC50XXS
90 seconds
360401S
105 seconds
380401S, 380402S, 384UV5S
180 seconds
360801S
135 seconds
380801S, 380802S, 88UV5S
210 seconds
To change the time, access the Menu, as described in step 13, step through the menu choices until CTO
is displayed, then press the ZERO button to select. Use the RANGE button to increment and the MODE
button to decrement the time. Pressing the ZERO and MODE buttons simultaneously will return to the
value last stored. Press ZERO to select the timeout factor and return to normal operation.
Values available are:
15 to 405 seconds in 15 second steps
12. Wall Mount Power Supply
Use the wall mount power supply to power the AI310 and charge the optional battery.
13. Optional Meter Mounting System
Scientech offers a sturdy mount system (Part Number MT) consists of a non-slip base with three
moldable feet, ball mounts, and a meter cradle. This system will hold your AI310 about 15cm above an
optical bench or work surface in an upright position. The ball mounts let you reposition your meter to
accommodate any viewing angle you desire. It also allows you to easily remove your meter at the end
of the day for safekeeping.