PN 10795T 15
To enter into the statistics mode:
A. Press the RANGE and MODE buttons simultaneously. The number of pulses in the last statistics
run and a flashing SET annunciator will appear in the display.
B. Use the RANGE (count up) and MODE (count down) buttons to change the display to the desired
number of pulses to include in the statistics run (up to 1000).
C. Press the SELECT button to enter the pulse population to memory. The display will blank and a
flashing STATS annunciator will appear.
D. Press the MODE button to begin the run. The STATS annunciator will stop flashing. The power
meter will automatically stop once the data has been collected.
D. Press the ON/SELECT button to recall the data to the display one item at a time. Each time the
ON/SELECT button is pressed the following statistical calculations will be sequentially displayed:
Number of pulses collected
Average energy (AVG)
Minimum energy (MIN)
Maximum energy (MAX)
Standard deviation (SIGMA)
Coefficient of variation (CV%)
F. Press the ON/SELECT button one more time to return to step D for a new stats run. The flashing
STATS annunciator indicates that the H410 is ready for a new run. Each time a new run begins
the data from the previous run is lost.
G. Press the OFF/CANCEL button to return the power meter to the mode of operation in effect prior
to statistics mode.
H. Press and hold the OFF/CANCEL until the display blanks to save the setting.
I. Press the ON/SELECT button to turn the H410 back on.
12. To Perform a Transfer Calibration