EasyManua.ls Logo

ScienTECH Vector Series - Page 6

ScienTECH Vector Series
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN 10795T 5
Table 1. H410 Specifications
Model
Handheld: H410
H410D
Display
4 Digit LCD with analog indicator
4 Digit backlight, LCD
Modes, Power or Energy
Pyroelectric: energy, average energy,
power, statistics
Calorimeter: energy, power, calibration
Pyroelectric: energy, average energy,
power, statistics
Calorimeter: energy, power, calibration
Pyroelectric Statistics Mode
Pulses collected, average energy, min
energy, max energy, standard deviation,
coefficient of variation
Pulses collected, average energy, min
energy, max energy, standard deviation,
coefficient of variation
For 25mm Calorimeters:
Power Full Scale Ranges
Energy Full Scale Ranges
10mW, 100mW, 1W, 10W, Auto
10mJ, 100mJ, 1J, 10J
10mW, 100mW, 1W, 10W, Auto
10mJ, 100mJ, 1J, 10J
For 50mm Calorimeters:
Power Full Scale Ranges
Energy Full Scale Ranges
300mW, 3W, 30W, Auto
300mJ, 3J, 30J
300mW, 3W, 30W, Auto
300mJ, 3J, 30J
For Pyroelectric Detectors:
Power Full Scale Ranges
Energy Full Scale Ranges
3mW, 30mW, 300mW, 3W, Auto
3mJ, 30mJ, 300mJ, 3J
3mW, 30mW, 300mW, 3W, Auto
3mJ, 30mJ, 300mJ, 3J
For HR Pyroelectric Detectors
Not supported
Not supported
For 100mm Calorimeter:
Use Interface Module, PN 10735
Use Interface Module, PN 10748
300mW, 3W, 30W, Auto or 300mJ, 3J,
30J
500mW, 5W, 50W, Auto or 1.5mJ, 15J,
150J
300mW, 3W, 30W, Auto or 300mJ, 3J,
30J
500mW, 5W, 50W, Auto or 1.5mJ, 15J,
150J
For 200mm Calorimeter:
Use Interface Module, PN 10747
Use Interface Module, PN 10769
300mW, 3W, 30W, Auto or 300mJ, 3J,
30J
1W, 1W, 100 W, Auto or 3J, 30J, 300J
300mW, 3W, 30W, Auto or 300mJ, 3J,
30J
1W, 1W, 100 W, Auto or 3J, 30J, 300J
Calorimeter max repetition rate,
power mode
Unlimited
Unlimited
Calorimeter max repetition rate,
energy mode
1 pulse every 60 to 90 sec, calorimeter
dependent
1 pulse every 60 to 90 sec, calorimeter
dependent
Pyroelectric, max repetition rate
collecting data in statistics mode
300 pps
300 pps
Response time for calorimeter,
power mode
3-10 seconds depending on calorimeter
3-10 seconds depending on calorimeter
Response time for calorimeter,
energy mode
1-3 seconds depending on calorimeter
1-3 seconds depending on calorimeter
Operating temperature
5 to 40
o
C
5 to 40
o
C
Power Requirements
120 Volts, 60 Hz ± 10 % or
220 Volts, 50 Hz ± 10 %
120 Volts, 60 Hz ± 10 % or
220 Volts, 50 Hz ± 10 %
Rechargeable Batteries
4 Ni-MH or NI-Cad, included
4 Ni-MH or NI-Cad, included
Dimensions D x L, inches
cm
8.25 x 4.0 x 1.5
20.96 x 10.16 x 3.81
8.25 x 4.0 x 1.5
20.96 x 10.16 x 3.81
Weight, lbs
kg
1.23
0.56
1.23
0.56
Other Specs:
Stand option: VHSTAND