11
3)ALHC(The alarm value of deviation of real temperature and the setting temperature, ℃, one decimal)
When real temperature is beyond this value, the equipment alarms continuously and the alarm indicator
turns on, heating output is automatically cut off. Press the “Increase”or “Decrease”button to set this
parameter, press the “Change/Confirm” button again to confirm the change.
Simplified instruction on the setting of temperature exceeding alarm
3.Setting of the clock
Press the “Change/Confirm” button to display the clock on the screen. The clock time can be changed
when deviation between clock and standard time occurs.
Hold on the “Change/Confirm” button for 3 seconds, screen displays “CIP”, a second later “0”, press the
“Increase”button to input the password “4”, press the “Change/Confirm” button again, screen displays
the hour value, press the “Increase”or “Decrease”button to change the hour value, press the
“Change/Confirm” button one more time, screen display the minute value, press the “Increase”or
“Decrease”button to change the minute value, press the “Change/Confirm” button at last to confirm the
change.
Simplified instruction on the setting of time
Attention!
1.When this ALHC value is “0”, equipment will not alarm.
2.Change the preset temperature or enlarge the ALHL value can terminate the
temperature exceeding alarm.
3.When deviation is back to the range you set, the equipment terminate the
temperature exceeding alarm automatically.
4.Press the “Operating”button can make the equipment stop running.