Technical data • 29
English
The two figures must match for calibration to be valid. If
the sticker and calibration counter do not match, recali-
bration is necessary. Contact your service partner or the
seca customer service department. Once the scale has
been recalibrated, a new, updated calibration sticker is
used to identify the calibration counter state.This sticker
is attached with an extra seal issued by the person
authorized to perform the recalibration. The calibration
counter sticker can be ordered fromseca customer
service.
10.TECHNICAL DATA
Technical data for seca 955
Dimensions
•Depth
•Width
•Height
956 mm
563 mm
897 mm
Weight approx. 25 kg
Temperature range +10° C to +40°C
Height of figures 25 mm
Power supply
Rechargeable battery block
Power pack
Power consumption approx. 32 mA
Maximum runtime in rechargeable battery
operation approx. 2.400 minutes
Medical device according to directive
93/42/EEC
Class I with measuring function
Calibration according to directive 2009/23/EC Class III
EN 60601-1:
Class II totally insulated appliance:
Type B electromedical device:
Maximum load
Partial weighing range 1
Partial weighing range 2
150 kg
250 kg
Minimum load
Partial weighing range 1
Partial weighing range 2
1.0 kg
2.0 kg