8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 23).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
9. Technical data
Maximum load 200 kg
Minimum load 1 kg
Graduations
seca 959:
0 up to 150 kg: 50 g
150 up to 200 kg: 100 g
seca 958:
0 up to 100 kg: 100 g
100 up to 200 kg: 200 g
Dimensions
height: 920 mm
width: 565 mm
depth: 810 mm
Weight approx. 24.5 kg
Temperature range +10 °C to +40 °C
Power supply rechargeable batteries/mains unit
Power consumption approx. 25 mA
Calibration as per Directive
90/384/EEC
seca 959: approved for medical use, Class III
seca 958: approved for medical use, Class IIII
Medical product as per Directive
93/42/EEC Class I
EN 60 601-1:
Insulated equipment, Protection Class II:
medical electrical equipment type B: