Specifies requirements for power line connection, voltage, and impedance.
Explains how DIP switches on the Portable Control Board define unit configuration.
Details the procedure for entering and saving configuration data via the control panel.
Describes how to select and configure Workstations 1 and 2.
Explains how to set mA stations for small and large focal spots.
Procedure for selecting the X-ray tube identifier.
Configures time settings for 'PREP' function and collimator light.
Sets the maximum percentage limit for X-ray tube power.
Details accessing and using service applications via the touch screen interface.
Steps for connecting the oscilloscope and preparing for calibration measurements.
Calibrates filament current for the lowest mA station in Large Focus at 40 and 80 kVp.
Procedure for calibrating the kVp Gain (P01) using a non-invasive kVp meter.
Procedure for calibrating the mA Gain (P02) before autocalibration.
Procedure for automatic calibration of mA stations for Large Focal Spot.
Manual calibration for filament current when autocalibration fails.
Steps to re-initialize the EEPROM memory on the Portable Control Board.
Instructions for updating the unit's software using an SD memory card.
Lists and explains various error codes encountered during operation.
Checks alignment of X-ray beam and field size indicators.
| Generator Type | High Frequency |
|---|---|
| Frequency | 50/60 Hz |
| Maximum Power | 8 kW |
| Power Output | 8 kW |
| Phases | Single |
| Cooling System | Forced Air |