3
Probe Calibration
Wash
Keep probe into Buffer solution
Press Cal Key 5 Seconds
Please Wait
Calibration During 1 minute
Quality Probe
Wash
Keep probe into Buffer solution
Press Enter Key and calibrate
second point 4 pH buffer solution
Please Wait during 1 minute
Quality Probe
Wash
Press Enter Key to save and exit
Normal Status
Note:
If you have setting EASY CAL the calibration function has 1 point calibrate only 7 pH buffer solution.