MANUAL DE INSTRUCCIONES CODIGO 80104 REV J 02/2021 Pag.: 11
J.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
e-mail: selecta@jpselecta.es - http://www.grupo-selecta.com
CONTROL PANEL:
- Select the temperature value with the Up/Down keys and on
screen 2 (in green), the value will ash.
TEMPERATURE CORRECTION:
You can correct the temperature displayed on the screen if
contrasted with a more accurate thermometer.
- Press the parameter input key to enter the setup menu.
- Press the parameter change key until the message iNS
appears.
- Adjust the temperature value with the Up / Down keys.
Example 1:
If the standard thermometer indicates 10°C while the MEDILOW
unit temperature control indicates 9 ºC. The value to be entered
is 1.
Example 2:
If the standard thermometer indicates 10ºC while the MEDILOW
unit temperature control indicates 11ºC. The value to be entered
is -1.
Note: For more details on regulator operation, use additional
specic instructions for it.
MAINTENANCE
Before removing the covers disconnect the equipment from
the mains.
The parameters of temperature regulator must not be
manipulated by unauthorized personnel. It can cause
damages of difcult solution.
When its moving is necessary, put the set vertically on a palet
whose dimensions (at least) are the same as the ones of the
base of the cabinet. Once the moving is nished, 12 hours
must be passed before starting up the set.
CLEANING:
The cleaning of the cabinet freezer frequently is recommended.
For it, disconnect the equipment from the power. Clean the
chamber and casing with lukewarm water with a small quantity
of detergent, rinse with clear water and dry well. Do not use
abrasive, alkaline, acid or base cleaning products.
DEFROSTING:
The equipment is equipped with an automatic defrost system.
However, when operating with the regulator in the coldest
settings, frost may occur on the inner bottom wall. In this case, the
equipment must be defrosted manually. To defrost it, disconnect
the equipment by removing the plug from the socket!

Other manuals for Selecta MEDILOW-S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Selecta MEDILOW-S and is the answer not in the manual?

Selecta MEDILOW-S Specifications

General IconGeneral
BrandSelecta
ModelMEDILOW-S
CategoryAccessories
LanguageEnglish