INSTRUCTION MANUAL CODE 80152 REV E DEC 2019 (They can be modied without notice) Page: 29
J.P. SELECTA s.a.u. Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
RS-232 INSTRUCTIONS FOR VACIOTEM T AND VACIOTEM TV OVENS
Vaciotem T / Vaciotem TV
PVT? Get probe temperature PV (Point value):
Command: #01PVT?KK<CR><LF>
Answer: #01+000KK<CR><LF>
Vaciotem TV
PVP? Get pressure value
Command: #01PVP?KK<CR><LF>
Answer: #010000KK<CR><LF>
Vaciotem T / Vaciotem TV
SVT? Get Set point temperature:
Command: #01SVT?KK<CR><LF>
Answer: #01+000KK<CR><LF>
SVT Set point temperature
Command: #01SVT<BL>+000KK<CR><LF>
Answer: #01OKKK<CR><LF>
Vaciotem TV
SVP? Get Set point temperature:
Command: #01SVP?KK<CR><LF>
Answer: #010000KK<CR><LF>
SVP Set point temperature
Command: #01SVP<BL>+0000KK<CR><LF>
Answer: #01OKKK<CR><LF>
Vaciotem T / Vaciotem TV
RUN? Get status variable
Command: #01RUN?KK<CR><LF>
Answer:
Temperature stop: #01STPTKK<CR><LF>
Pressure stop: #01STPPKK<CR><LF>
Temperature run: #01RUNTKK<CR><LF>
Pressure run: #01RUNPKK<CR><LF>
Alarm status: #01ALARKK<CR><LF>
Standby time: #01STBYKK<CR><LF>
Model Vaciotem TV gives compound answers. For exemple:
#01STPT<BL>STPPKK<CR><LF>
Vaciotem TV
RUN Starts run:
Command: #01RUNKK<CR><LF>
Answer: #01OKKK<CR><LF>
Vaciotem T / Vaciotem TV
RUNT Starts temperature control:
Command: #01RUNTKK<CR><LF>
Answer: #01OKKK<CR><LF>