MANUAL DE INSTRUCCIONES CODIGO 80152 REV E DIC 2019 (Sujetas a modicaciones sin previo aviso) Pag.: 30
J.P. SELECTA s.a.u. Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
Vaciotem TV
RUNP Starts pressure control:
Command: #01RUNPKK<CR><LF>
Answer: #01OKKK<CR><LF>
Vaciotem TV
STOP Stops device:
Command : #01STOPKK<CR><LF>
Answer : #01OKKK<CR><LF>
Vaciotem T / Vaciotem TV
STPT Stops temperature control:
Command : #01STPTKK<CR><LF>
Answer : #01OKKK<CR><LF>
Vaciotem TV
STPP Stops device:
Command : #01STPPKK<CR><LF>
Answer : #01OKKK<CR><LF>
Vaciotem T / Vaciotem TV
ID? Getsdeviceidentication:
(Software version and device code)
Command : #01ID?KK<CR><LF>
Answer : #014001489V01KK<CR><LF> (Vaciotem-T)
#014001490V01KK<CR><LF> (Vaciotem-TV)
7 characters for code + V + 2 for the version
Vaciotem T / Vaciotem TV
SAL? Gets alarm status:
Command : #01SAL?KK<CR><LF>
Answer : #01ALARM1KK<CR><LF>
* See table of the alarm status.
Vaciotem T / Vaciotem TV
TBY? Gets standby time:
Command : #01TBY?KK<CR><LF>
Answer : #0100,0KK<CR><LF>
SBY Sets standby time:
Command : #01TBY?KK<CR><LF>
Answer : #0100,0KK<CR><LF>
Time in hours and tenth parts of an hour
Between 0 and 24 hours