INSTRUCTION MANUAL CODE 80152 REV E DEC 2019 (They can be modied without notice) Page: 31
J.P. SELECTA s.a.u. Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
TBR? Gets standby set time:
Command : #01TBR?KK<CR><LF>
Answer : #0100,0KK<CR><LF>
Vaciotem T / Vaciotem TV
TRU? Gets run time:
Command : #01TRU?KK<CR><LF>
Answer : #010,00KK<CR><LF>
Answer : #0100,0KK<CR><LF>
Time until 9h59mim (9,59)
Time from 10h until 9.99 (99,9)
TRU Sets run time:
Command : #01TRU<BL>0,00KK<CR><LF>
Answer : #01TRU<BL>00,0KK<CR><LF>
Answer : #01OKKK<CR><LF>
Time until 9h59mim (9,59)
Time from 10h until 9.99 (99,9)
Vaciotem TV
INT Inhibits/enable key pad (temperature/pressure):
Command : #01INT<BL>1KK<CR><LF>
Answer : #01OKKK<CR><LF>
1 = inhibit 0 = enable
Vaciotem T/Vaciotem TV
ITT Inhibits/enable key pad (temperature/pressure):
Command : #01INT<BL>1KK<CR><LF>
Answer : #01OKKK<CR><LF>
1 = inhibit 0 = enable
Vaciotem TV
ITP Inibits/enable key pad (temperature/pressure):
Command : #01INT<BL>1KK<CR><LF>
Answer : #01OKKK<CR><LF>
1 = inhibit 0 = enable
Vaciotem TV
SRLK Opens pump valve :
Command : #01SRLKKK<CR><LF>
Answer : #01OKKK<CR><LF>
RRLK Closes pump valve :
Command : #01RRLKKK<CR><LF>
Answer : #01OKKK<CR><LF>